中文 | English

Providing a Total Solution for all your X-Ray System needs

Technical Parameters
Gemini 0150F10
Basic 
Output Power  150 W  
Tube Voltage Range (kV)  25-100
Tube Current Range (mA)  0.05-6 
Mechanical 
Tank Dimension (mm)  308 × 205 × 150
Tank Weight (kg)  17.2
Inverter Dimension (mm)  301 × 152 × 55 
Inverter Weight (kg)  1.5
Electrical 
Tube Voltage Ripple  <0.5% (RMS)                                                                                           
Tube Voltage Accuracy  ± 1%
Tube Voltage Rising Time (ms)  500
Tube Current Accuracy  ± 1%
Power Supply
Mains Input 

 1Φ, 90~264VAC / 50-60 Hz

Cooling Fan Input

1Φ,220VAC or 110VAC/50-60Hz

X Ray Tube / Tank 
Focal Spot Size (mm)  0.5 
Beam Geometry  80° ×  20° Fan                                                                                          
Operational                                                                                     
Interface  RS232                                                                   
Software  GUI service tool                                                                                        
Operating Mode  Continuous Exposure Mode                                                   
Thermal    
Max. Tank Heat Storage  450 kHU / 338 kJ                                                                                 
Tank Heat Dissipation (W)  150                                                                                
Environmental 
Operating Temperature (℃)  0-40
Storage & Transportation Temperature (℃)  -20-70
Altitude  0 to 4000m
Operating Humidity(Non-condensing) 5-95
Storage & Transportation Humidity (Non-condensing) 5-95

 

 

 

 

 

 

* This document cannot be considered to be a contractual specification.

The information given herein may be modified without notice.