中文 | English

Providing a Total Solution for all your X-Ray System needs

Technical Parameters
Norma 201i
Chamber Ionization Chamber Input Voltage ±12 V or ± 15 V ± 5% (DC) 
Number of Measuring Fields 2 Input Current (mA)
≤ 200 @+12V, ≤100 @-12V
Energy Range (kV) 40 ~ 150 Dimensions (mm) 475 × 460 × 3.3
Exposure Time Range 1ms ~ 10s Weight (kg) 1.3
Sensitivity Typ. at 80 kV, 21 mm Al
0.2 V/μGy; Adjustable
Operating Temperature () 10 ~ 40
Dose Range (µGy) 1 ~ 50 Storage & Transport Temperature with Package () -40 ~ 70
Sensitivity Tolerance 3% @ Default Sensitivity Operating Humidity (% RH)  10 ~ 90 (Non-Condensing)
Sensitivity Tolerance Between Measure Fields 5% @ Default Sensitivity Storage & Transport Humidity with Package (% RH)  10 ~ 90 (Non-Condensing)
X-ray Transmittance 85% @ 40 kV
92% @ 80 kV
Operating Atmospheric Pressure (mbar) 700 ~ 1060
Interfield Voltage in Ionization Chamber (V) 102 ± 10    

 

* This document cannot be considered to be a contractual specification.

The information given herein may be modified without notice.

Dimensions